Description

Woson is a renowned manufacturer of high-quality medical and cosmetic equipment. The brand supplies professional equipment to beauty salons , nail salons , tattoo parlors , hospitals, and clinics – equipment that plays a crucial role in the sterilization process and thus contributes to a maximum level of safety and hygiene for both practitioners and clients.

Safe sterilization for a wide range of instruments
The compact Woson New Tanco 12 L autoclave is indispensable in any environment where sterile tools make all the difference: from beauty and nail salons to medical practices and studios. Sterilization with a top-of-the-line medical class B autoclave offers complete effectiveness and process safety , allowing you to consistently maintain high hygiene standards.
The device is designed for sterilizing both medical and non-medical instruments: packaged and unpackaged , solid , hollow (type A) , and porous . Caution: Operation is intended only for qualified personnel with knowledge and appropriate training for this type of equipment.

Improvements for additional functionality and control
The autoclave features a control panel with an LED display that allows you to easily select programs and parameters and accurately monitor the process. The built-in thermal printer allows you to print a sterilization certificate with information such as the date and time , temperature curve , and pressure . There's also a connection on the back for connecting a dedicated additional printer, if needed.
The USB port also allows you to save sterilization results digitally on the included memory stick. Built-in tanks for distilled and used water accelerate and simplify daily work. The appliance features water quality monitoring . The self-diagnosis system provides clear messages in the event of errors (e.g., door not properly closed, empty tank, or replace water), reducing the risk of damage to the appliance.

Available sterilization, testing and cleaning programs
The autoclave is equipped with: 4 sterilization programs , 2 test programs (allowing you to perform three tests: Vacuum Test , B&D Test , and Helix Test ), a pipe maintenance program ( Automatic Pipe Cleaning ), and a drying program for higher drying requirements ( Drying Function ). You can customize the parameters to your load and working methods, and save your own settings via the personalized mode ( CUSTOM ).
STERILIZATION PROGRAMS
| Sterilization program
|
Type of load
|
Temperature
|
Busy
|
Number of vacuum cycles
|
Cycle time
|
| Unpacked |
Unwrapped Item Sterilization Program |
134°C |
210 kPa |
1x |
Sterilization: 4 min; drying: 9 min |
| PACKED |
Packaged Item Sterilization Program |
134°C |
210 kPa |
3x |
Sterilization: 5 min; drying: 9 min |
| PRION |
Program for inactivation of prions, bacteria and viruses |
134°C |
210 kPa |
3x |
Sterilization: 18 min; drying: 9 min |
| POROUS |
Program for porous items such as cotton and similar materials |
121°C |
110 kPa |
3x |
Sterilization: 20 min; drying: 18 min |

Specification
| Medical class B
|
The most advanced sterilizers for effective sterilization of packaged and unpackaged items, with fractional pre-vacuum. |
| Regulations & certification
|
CE 0197; complies with Directive 93/42/EEC and 97/23/EC. Conforms to EN 61010-1, EN 61010-2-040, EN 13060, EN 61326-1. Manufacturer has EN ISO 9001 and EN ISO 13485 certification. |
| Capacity 12 liters
|
Ideal for salons and clinics. Suitable for a basket with three trays for packaged and unpackaged tools. |
| Built-in printer
|
Works with thermal paper, without the hassle of replacing ink. |
| Sterilization packaged and unpackaged
|
High effectiveness for both sterilization methods. |
| LED display |
For easy program selection and parameter setting; shows temperature, pressure and time during the cycle. |
| 8 automatic programs
|
4 sterilization programs, 2 test programs, 1 pipe cleaning program and 1 drying program. |
| Tanks for clean and used water
|
Fresh water tank for distilled/deionized water; used water tank eliminates the need for an external container. |
| Self-diagnosis system
|
Notifications help to identify errors in time and prevent damage. |
| Stainless steel chamber
|
Durable, medically suitable material that underlines the high quality of workmanship. |
| Sanitary & epidemiological
|
Complies with international hygiene standards. |
| Technical passport
|
Included in the set. |
| Warranty
|
24 months; warranty period does not depend on the number of cycles performed. |

Technical data
| Certificates and standards
|
CE 0197 (93/42/EEC, 97/23/WE) EN 61010-1, EN 61010-2-040, EN 13060, EN 61326-1 |
| Assets
|
1600 W |
| Nominal voltage
|
220VAC ~ 240VAC, 50Hz~60 Hz |
| Maximum operating temperature
|
145°C |
| Maximum operating pressure
|
2.5 bar (250 kPa) |
| Minimum operating pressure
|
-0.9 bar (90 kPa) |
| Time setting
|
121°C: 20–60 min; 134°C: 4–20 min |
| Drying time
|
0–60 min (step size 5 min) |
| Vacuum cycles
|
1x, 3x or 5x |
| Package dimensions
|
652 x 549 x 486 mm |
| Device dimensions (external)
|
545 x 382 x 375 mm |
| Material room
|
RVS (stal nierdzewna) |
| Room dimensions
|
Depth 295 mm ø 190 mm |
| Internal tray dimensions
|
245 x 145 x 15 mm |

Read the manual carefully before use.
Included: autoclave, tray frame, three trays, handle for removing the tray, hose for draining used water, power cord, door seal, user manual, technical passport and pen drive.
Frequently Asked Questions
1) Which locations is the Woson New Tanco 12 L suitable for?
For beauty salons , nail studios , tattoo studios , clinics and practices where reliable sterilization is essential.
2) What does medical class B mean for an autoclave?
Class B is the highest sterilization class and is suitable for packaged and unpackaged loads, including hollow and porous items.
3) Can I keep sterilization certificates for administrative purposes?
Yes. You can print results using the built-in thermal printer and also save them digitally via the USB port on the included memory stick.
4) What programs are on this device?
A total of 8 automatic programs : 4 sterilization programs, 2 test programs, 1 pipe cleaning program and 1 drying program.
5) What tests can I perform?
Vacuum Test, B&D Test and Helix Test (via the available test programs).
6) What is the advantage of built-in water tanks?
You don't need any external containment: there is a tank for clean distilled/deionized water and a tank for used water , which speeds up the workflow.
7) How does the self-diagnosis system help in practice?
The device displays clear messages (such as door not closed properly, empty tank or replace water), allowing you to resolve errors more quickly and prevent damage .
8) Is this autoclave suitable for packaged instruments?
Yes, the device is designed for packaged and unpackaged instruments and offers high efficacy for both.
9) Can anyone operate this device?
No. According to the directive, the device may only be operated by qualified personnel with appropriate training.
10) What about warranty?
The warranty is 24 months and is not dependent on the number of cycles performed.